Actions
atm-environment-alarm-system #29
closed
CP
SPS30 Maintenance (for the Manual)
atm-environment-alarm-system #29:
SPS30 Maintenance (for the Manual)
Status:
Closed
Priority:
normal
Assignee:
-
% Done:
0%
Type:
Implementation
Description
Maintenance¶
The SPS30 is designed to be maintenance-free due to its "sheath air" technology, which protects the optical components from fouling.
Automatic Fan Cleaning:¶
- The sensor features an internal cleaning routine that spins the fan to maximum speed for 10 seconds to blow out accumulated dust.
- Default Interval: Every 168 hours (1 week) of continuous operation.
- Power Cycles: If the sensor is turned off and on frequently, the internal timer resets to zero. In this case, you should manually trigger a cleaning cycle at least once a week.
Manual Cleaning: Manual cleaning is generally not required and should not involve opening the casing unless it has failed, as this can damage the internal HEPA filter or optical alignment.¶
CP Updated by Carsten Poulsen 4 months ago
The SPS30 will be permanently powered, so it should automatically clean once a week.
CP Updated by Carsten Poulsen 23 days ago
- Status changed from New to Ongoing
CA Updated by claude assistant 23 days ago
- Status changed from Ongoing to Done
CA Updated by claude assistant 23 days ago
Added a 'Maintenance' section to Particle Sensor User Manual.docx (before Technical Specifications), covering the SPS30's maintenance-free sheath-air design, its automatic weekly fan-cleaning cycle, and a caution against opening the casing for manual cleaning. Committed as f1dba02 and pushed to origin/master. Status set to Done.
CP Updated by Carsten Poulsen 23 days ago
- Status changed from Done to Closed